Shape-based pose estimation for automatic standard views of the knee

05/26/2023
by   Lisa Kausch, et al.
0

Surgical treatment of complicated knee fractures is guided by real-time imaging using a mobile C-arm. Immediate and continuous control is achieved via 2D anatomy-specific standard views that correspond to a specific C-arm pose relative to the patient positioning, which is currently determined manually, following a trial-and-error approach at the cost of time and radiation dose. The characteristics of the standard views of the knee suggests that the shape information of individual bones could guide an automatic positioning procedure, reducing time and the amount of unnecessary radiation during C-arm positioning. To fully automate the C-arm positioning task during knee surgeries, we propose a complete framework that enables (1) automatic laterality and standard view classification and (2) automatic shape-based pose regression toward the desired standard view based on a single initial X-ray. A suitable shape representation is proposed to incorporate semantic information into the pose regression pipeline. The pipeline is designed to handle two distinct standard views simultaneously. Experiments were conducted to assess the performance of the proposed system on 3528 synthetic and 1386 real X-rays for the a.-p. and lateral standard. The view/laterality classificator resulted in an accuracy of 100%/98% on the simulated and 99%/98% on the real X-rays. The pose regression performance was dθ_a.-p=5.8±3.3, dθ_lateral=3.7±2.0 on the simulated data and dθ_a.-p=7.4±5.0, dθ_lateral=8.4±5.4 on the real data outperforming intensity-based pose regression.

READ FULL TEXT

page 2

page 4

page 8

research
06/22/2018

Augmented Reality-based Feedback for Technician-in-the-loop C-arm Repositioning

Interventional C-arm imaging is crucial to percutaneous orthopedic proce...
research
07/11/2022

SDFEst: Categorical Pose and Shape Estimation of Objects from RGB-D using Signed Distance Fields

Rich geometric understanding of the world is an important component of m...
research
09/22/2017

Real-time 3D Shape Instantiation from Single Fluoroscopy Projection for Fenestrated Stent Graft Deployment

Robot-assisted deployment of fenestrated stent grafts in Fenestrated End...
research
12/11/2020

A novel joint points and silhouette-based method to estimate 3D human pose and shape

This paper presents a novel method for 3D human pose and shape estimatio...
research
11/06/2022

Towards real-time 6D pose estimation of objects in single-view cone-beam X-ray

Deep learning-based pose estimation algorithms can successfully estimate...
research
03/22/2021

3D Reconstruction and Alignment by Consumer RGB-D Sensors and Fiducial Planar Markers for Patient Positioning in Radiation Therapy

BACKGROUND AND OBJECTIVE: Patient positioning is a crucial step in radia...
research
11/29/2014

Egocentric Pose Recognition in Four Lines of Code

We tackle the problem of estimating the 3D pose of an individual's upper...

Please sign up or login with your details